site stats

Csi wirion recall

WebNov 26, 2024 · Benzinga. Nov. 26, 2024, 07:55 AM. Cardiovascular Systems Inc (NASDAQ:CSII) has commenced a voluntary recall of its WIRION Embolic Protection … WebMar 14, 2024 · WIRION ® is indicated for use as an embolic protection system to contain and remove embolic material (thrombus/debris) while performing atherectomy in calcified …

WIRION Indication & Safety Information CSI360

WebCSI representative or call customer service at 1-877-274-0901. PACKAGE INCLUDES: (1) Delivery Catheter with Pre-Loaded Filter, (1) Retrieval Catheter (1) Syringe with Flushing Tip (5 cc) WIRION® is indicated for use as an embolic protection system to contain and remove embolic material (thrombus/debris) while performing atherectomy WebMar 17, 2024 · Device offers a streamlined approach for lower extremity embolic protection Cardiovascular Systems, Inc. (CSI ® ) (NASDAQ: CSII), a medical device company … how many children does glenn beck have https://hellosailortmh.com

Cardiovascular Systems, Inc. Acquires Gardia Medical Ltd.

WebNov 25, 2024 · Medical device company Cardiovascular Systems (CSI) has commenced a voluntary recall of its WIRION Embolic Protection Systems (WIRION). The unused WRN … WebJul 6, 2024 · CSI, based in St. Paul, Minnesota, sent an urgent recall notice to all affected customers asking them to discontinue use of all Wirion units and return unused product … WebNov 22, 2024 · Recall Status 1: Completed: Recall Number: Z-0394-2024: Recall Event ID: 89120: 510(K)Number: K210282 K213372 Product Classification: Temporary Carotid … high school in midland tx

WIRION Indication & Safety Information CSI360

Category:Indications Statements, Important Safety & Risk Information

Tags:Csi wirion recall

Csi wirion recall

Class 1 Device Recall WIRION EMBOLIC PROTECTION SYSTEM

WebNov 24, 2024 · The company plans to voluntarily recall all WIRION units currently in customer inventory. In total, 697 devices were distributed in the U.S. between March 22, … WebNov 22, 2024 · Action. Consignees were issued an URGENT: Medical Device Recall letter dated 11/22/2024. Consignees are asked to immediately remove from use and return affected devices to CSI. To do so: 1) Complete the attached Customer Acknowledgement Form. a) Per lot, document the quantity of all devices that are to be returned.

Csi wirion recall

Did you know?

WebMar 17, 2024 · The diameter of the vessel at the site of filter basket placement should be between 3.5 to 6 mm. Wirion may be used with commercially available 0.014-inch guidewires, advised the company. In August 2024, CSI acquired the Wirion device and related assets from Gardia Medical Ltd., a wholly owned Israeli subsidiary of Allium … WebNov 25, 2024 · Cardiovascular Systems has initiated a voluntary recall of unused Wirion embolic protection systems due to complaints of filter breakage during retrieval. Wirion is a distal embolic protection filter used …

WebNov 25, 2024 · CSI recalls Wirion embolic protection device. 25th November 2024. 3322. Cardiovascular Systems has initiated a voluntary recall of unused Wirion embolic … WebNov 25, 2024 · Cardiovascular Systems has initiated a voluntary recall of unused Wirion embolic protection systems due to complaints of filter breakage during retrieval. Wirion …

Web2 days ago · This product is distributed by CSI. Manufactured By. WIRION ® Peripheral Embolic Protection System. Indications Statement: WIRION® is indicated for use as an embolic protection system to contain and remove embolic material (thrombus/debris) while performing atherectomy in calcified lesions of the lower extremities. The diameter of the … WebSt. Paul, Minn. – November 24, 2024 – Cardiovascular Systems, Inc. (CSI ... The company plans to voluntarily recall all WIRION units currently in customer inventory. In total, 697 …

WebNov 25, 2024 · CSI recalls Wirion embolic protection device. 25th November 2024. 3322. Cardiovascular Systems has initiated a voluntary recall of unused Wirion embolic protection systems due to complaints of filter breakage during retrieval. Wirion is a distal embolic protection filter used to capture thrombus and debris that can be associated with all types ...

WebNov 24, 2024 · The company plans to voluntarily recall all WIRION units currently in customer inventory. In total, 697 devices were distributed in the U.S. between March 22, 2024 and November 15, 2024. how many children does gogo skhotheni haveWebNov 24, 2024 · November 24, 2024—Cardiovascular Systems, Inc. (CSI) announced it has initiated a voluntary recall of unused Wirion embolic protection systems because of … how many children does giuliana rancic haveThe WIRION® system is used to hold and remove debris or blood clots from the lower limbs that can be associated with atherectomyExternal Link Disclaimerprocedures. See more Customers in the U.S. with questions about this recall should contact Cardiovascular Systems Inc. by phone at 651-259-2800. See more Cardiovascular Systems Inc. is recalling the WIRION system due to complaints of filter breakage during retrieval. Under certain … See more On November 22, 2024, Cardiovascular Systems Inc. sent an urgent notification recall letter to customers instructing them: 1. Remove the device from distribution and return the device to Cardiovascular Systems Inc. See more high school in my zoneWebNov 26, 2024 · CSI expects to introduce the acquired catheters as its new ViperCross line in the first half of its fiscal year, ending on 30 June 2024. Price Action: CSII shares closed … high school in my areaWebNov 24, 2024 · CSI has received nine complaints of filter breakage during retrieval. Filter breakage may result in device embolism and possible additional intervention; The company plans to voluntarily recall all Wirion units currently in customer inventory. In total, 697 devices were distributed in the U.S. this year between March 22 and Nov. 15 high school in munich germanyWebNov 26, 2024 · CSI has also notified the US Food and Drug Administration (FDA) regarding the voluntary recall. WIRION is a rapid exchange, pre-loaded filter, used as an embolic … high school in muntinlupa cityWebNov 26, 2024 · Print. Cardiovascular Systems Inc. announced a voluntary recall of its Wirion device on Wednesday afternoon. The company has told health care facilities to … high school in nc